CARELINK PROGRAMMER 2090

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-31 for CARELINK PROGRAMMER 2090 manufactured by Medtronic, Inc..

Event Text Entries

[187206753] Analysis confirmed the customer's stylus/touch screen issue. The connector between the xy board and overlay was cleaned and re-seated and the stylus was re-calibrated. The hard drive was reconfigured and the software was reloaded and updated. All found defective parts were replaced and all other identified issues were resolved. The device passed all final functional tests. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[187206754] It was reported that there was a display issue with the programmer. The pen had an erratic touch on screen. A second pen was used but the issue was unresolved. There was no patient involvement. It was further reported that the programmer was returned for service.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2182208-2020-00624
MDR Report Key9902361
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-03
Date Mfgr Received2020-03-03
Device Manufacturer Date2011-06-29
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1MEDTRONIC, INC.
Manufacturer Street8200 CORAL SEA STREET NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal Code55112
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARELINK PROGRAMMER
Generic NamePROGRAMMER, PACEMAKER
Product CodeKRG
Date Received2020-03-31
Returned To Mfg2020-03-09
Model Number2090
Catalog Number2090
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC, INC.
Manufacturer Address8200 CORAL SEA STREET NE MOUNDS VIEW MN 55112 US 55112


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.