EVERA MRI XT VR SURESCAN DVMB2D4

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-03-31 for EVERA MRI XT VR SURESCAN DVMB2D4 manufactured by Medtronic Europe Sarl.

Event Text Entries

[185978521] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[185978522] It was reported that it was suspected the patient received inappropriate shocks for supra ventricular tachycardia (svt) that the implantable cardioverter defibrillator (icd) detected as ventricular tachycardia (vt). It was noted that the icd detection discriminator showed some noise. The icd remains in use. No further patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9614453-2020-01117
MDR Report Key9902537
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-13
Date Mfgr Received2020-03-16
Device Manufacturer Date2019-05-10
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1MEDTRONIC EUROPE SARL
Manufacturer StreetROUTE DU MOLLIAU 31 CASE POSTALE
Manufacturer CityTOLOCHENAZ VAUD 1131
Manufacturer CountryCH
Manufacturer Postal Code1131
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEVERA MRI XT VR SURESCAN
Generic NameDEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Product CodeLWS
Date Received2020-03-31
Model NumberDVMB2D4
Catalog NumberDVMB2D4
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC EUROPE SARL
Manufacturer AddressROUTE DU MOLLIAU 31 CASE POSTALE TOLOCHENAZ VAUD 1131 CH 1131


Patients

Patient NumberTreatmentOutcomeDate
101. Life Threatening 2020-03-31