MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-31 for PRGM BI75000022 OARM ADDITIONAL manufactured by Medtronic Navigation, Inc (littleton).
[185767453]
Patient information unavailable at the time of filing. A manufacturer representative went to the site to test the imaging system. The manufacturer representative found blown 600v 6a fuse on stand alone board. Replaced fuse and performed system checkout with no more issues. System is functioning as intended so returned to service. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[185767514]
Medtronic received information regarding an imaging system used during a sacroiliac and thoracolumbar procedure. It was reported that the site could not shoot x-ray and that the system was stuck in initializing please wait. Multiple reboots no fix. Used the o2 for the case. There was a delay of less than 1 hour. No patient impact was correlated with this event.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3004785967-2020-00428 |
MDR Report Key | 9902598 |
Report Source | COMPANY REPRESENTATIVE,HEALTH |
Date Received | 2020-03-31 |
Date of Report | 2020-03-31 |
Date of Event | 2020-03-11 |
Date Mfgr Received | 2020-03-11 |
Device Manufacturer Date | 2015-04-17 |
Date Added to Maude | 2020-03-31 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | STACY RUEMPING |
Manufacturer Street | 7000 CENTRAL AVENUE NE RCW215 |
Manufacturer City | MINNEAPOLIS MN 55432 |
Manufacturer Country | US |
Manufacturer Postal | 55432 |
Manufacturer Phone | 7635260594 |
Manufacturer G1 | MEDTRONIC NAVIGATION, INC (LITTLETON) |
Manufacturer Street | 300 FOSTER ST |
Manufacturer City | LITTLETON MA 01460 |
Manufacturer Country | US |
Manufacturer Postal Code | 01460 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | PRGM BI75000022 OARM ADDITIONAL |
Generic Name | IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE |
Product Code | OXO |
Date Received | 2020-03-31 |
Model Number | BI75000022 |
Catalog Number | BI75000022 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MEDTRONIC NAVIGATION, INC (LITTLETON) |
Manufacturer Address | 300 FOSTER ST LITTLETON MA 01460 US 01460 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-31 |