SPRINT QUATTRO SECURE MRI SURESCAN 6947M62

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-31 for SPRINT QUATTRO SECURE MRI SURESCAN 6947M62 manufactured by Mpri.

Event Text Entries

[185739784] Concomitant medical products: model: dtba1d4, cardiac resynchronization therapy defibrillator (crt-d); implant: (b)(6) 2017. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[185739785] It was reported that oversensing was observed on the right ventricular (rv) lead on stored electrograms (egm). The lead remains in use. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2649622-2020-06652
MDR Report Key9902753
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2018-07-23
Date Mfgr Received2020-03-11
Device Manufacturer Date2012-04-03
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1MPRI
Manufacturer StreetROAD 149 KM 56.3
Manufacturer CityVILLALBA PR 00766
Manufacturer CountryUS
Manufacturer Postal Code00766
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSPRINT QUATTRO SECURE MRI SURESCAN
Generic NameDRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES
Product CodeNVY
Date Received2020-03-31
Model Number6947M62
Catalog Number6947M62
Device Expiration Date2014-03-23
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMPRI
Manufacturer AddressROAD 149 KM 56.3 VILLALBA PR 00766 US 00766


Patients

Patient NumberTreatmentOutcomeDate
15076 2020-03-31

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.