MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-31 for CARELINK PROGRAMMER 2090 manufactured by Medtronic, Inc..
[187206507]
Product analysis:the customer comment was confirmed at analysis. There was no backlight and the image on the screen was very weak. The flex cable display was damaged. A sharp fold caused the errors. The paper drawer was nearly empty, the latch media bay door was broken, the manufacturer logo button was damaged, the stylus was defective. The flex cable display was replaced, paper was added, the media bay door was replaced, the manufacturer logo button was replaced, the stylus was replaced, the touchscreen was calibrated, software was re-installed and updated to the newest version. All found defective parts were replaced and all other identified issues were resolved. The device then passed all final functional tests. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[187206508]
It was reported that the programmer display was very blurry and not visible. The programmer was returned for service. There was no patient involvement. It was further reported that the programmer subsequently tested out of specification during manufacturer's analysis.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2182208-2020-00628 |
MDR Report Key | 9902838 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-03-31 |
Date of Report | 2020-03-31 |
Date of Event | 2020-01-29 |
Date Mfgr Received | 2020-03-26 |
Device Manufacturer Date | 2011-06-21 |
Date Added to Maude | 2020-03-31 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | PAULA BIXBY |
Manufacturer Street | 8200 CORAL SEA ST NE |
Manufacturer City | MOUNDS VIEW MN 55112 |
Manufacturer Country | US |
Manufacturer Postal | 55112 |
Manufacturer Phone | 7635055378 |
Manufacturer G1 | MEDTRONIC, INC. |
Manufacturer Street | 8200 CORAL SEA STREET NE |
Manufacturer City | MOUNDS VIEW MN 55112 |
Manufacturer Country | US |
Manufacturer Postal Code | 55112 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CARELINK PROGRAMMER |
Generic Name | PROGRAMMER, PACEMAKER |
Product Code | KRG |
Date Received | 2020-03-31 |
Returned To Mfg | 2020-01-30 |
Model Number | 2090 |
Catalog Number | 2090 |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MEDTRONIC, INC. |
Manufacturer Address | 8200 CORAL SEA STREET NE MOUNDS VIEW MN 55112 US 55112 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-31 |