2CM PERIPHERAL CUTTING BALLOON 24628

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-31 for 2CM PERIPHERAL CUTTING BALLOON 24628 manufactured by Boston Scientific Corporation.

Event Text Entries

[185944168] (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[185944169] It was reported that a blade was missing. The 75% stenosed target lesion was located in the moderately tortuous cephalic vein. A 6. 00mm/ 2. 0cm/ 50cm peripheral cutting balloon was selected for use. During procedure, dilatation was performed twice at 10 atm around the cephalic vein anastomosis (the angle was gently/slowly applied). The device was simply removed since the lesion area was properly expanded. However, one out of the four blade was missing even if excessive force more than usual was not used. The procedure was completed with this device. There were no complications reported and there was no problem with the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-04092
MDR Report Key9903123
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-09
Date Mfgr Received2020-03-09
Device Manufacturer Date2019-12-06
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAY JOHNSON
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634942574
Manufacturer G1BOSTON SCIENTIFIC IRELAND LIMITED
Manufacturer StreetBALLYBRIT BUSINESS PARK
Manufacturer CityGALWAY
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name2CM PERIPHERAL CUTTING BALLOON
Generic NameCATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Product CodeLIT
Date Received2020-03-31
Returned To Mfg2020-03-25
Model Number24628
Catalog Number24628
Lot Number0024884443
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer AddressTWO SCIMED PLACE MAPLE GROVE MN 55311 US 55311


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31

© 2026 FDA.report