EPIC STENTED PORCINE HEART VALVE W/FLEXFIT SYSTEM E100-27M

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-31 for EPIC STENTED PORCINE HEART VALVE W/FLEXFIT SYSTEM E100-27M manufactured by St. Jude Medical, Brasil Ltda. (pampulha).

Event Text Entries

[187626603] The results/method and conclusion codes along with investigation results will be provided in the final report.
Patient Sequence No: 1, Text Type: N, H10


[187626604] On (b)(6) 2014, a 27mm epic valve was implanted with everting mattress suture technique using pledgets in the patient's mitral position in another hospital due to rheumatic mitral insufficiency. Early 2020, the patient was re-hospitalized due to mitral regurgitation and aortic regurgitation. A double valve replacement (dvr) was performed on (b)(6) 2020. The epic valve was explanted and replaced with a 25mm ats open pivot bileaflet heart valve. Upon explant, a tear was observed along the commissure between leaflets. No infection or pannus in growth was observed on the explanted valve.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3001883144-2020-00029
MDR Report Key9903133
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-02
Date Mfgr Received2020-03-04
Device Manufacturer Date2014-05-22
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAMELA YIP
Manufacturer Street5050 NATHAN LANE N
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1ST. JUDE MEDICAL, BRASIL LTDA. (PAMPULHA)
Manufacturer StreetRUA PROFESSOR JOSE VIEIRA DE MENDON ENGENHO NOGUEIRA - BELO HORIZONTE - MG
Manufacturer CityBELO HORIZONTE 31310-260
Manufacturer CountryBR
Manufacturer Postal Code31310-260
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEPIC STENTED PORCINE HEART VALVE W/FLEXFIT SYSTEM
Generic NameHEART-VALVE, NON-ALLOGRAFT TISSUE
Product CodeLWR
Date Received2020-03-31
Returned To Mfg2020-03-23
Model NumberE100-27M
Lot Number4652026
Device Expiration Date2018-05-21
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL, BRASIL LTDA. (PAMPULHA)
Manufacturer AddressRUA PROFESSOR JOSE VIEIRA DE MENDON?A 1301 ENGENHO NOGUEIRA - BELO HORIZONTE - MG BELO HORIZONTE 31310-260 BR 31310-260


Patients

Patient NumberTreatmentOutcomeDate
101. Other; 2. Required No Informationntervention 2020-03-31

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.