SPRINT 6945-65

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-31 for SPRINT 6945-65 manufactured by Mpri.

Event Text Entries

[185709224] Concomitant medical products: 419478 lead, implanted: (b)(6) 2005. Dtba1d1 crt-d, implanted: (b)(6) 2014. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[185709225] It was reported that the right ventricular (rv) lead exhibited unexplained counts to the sensing integrity counter (sic) as well as suspected oversensing of atrial fibrillation (af). Additionally there have been high rate non-sustained tachycardia. The rv lead remains in use. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


[188941002] Product analysis summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed. Analysis of the device memory indicated oversensing due to non-physiologic signals/sensing integrity counter. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2649622-2020-06677
MDR Report Key9903170
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-10
Date Mfgr Received2020-03-10
Device Manufacturer Date2001-02-20
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1MPRI
Manufacturer StreetROAD 149 KM 56.3
Manufacturer CityVILLALBA PR 00766
Manufacturer CountryUS
Manufacturer Postal Code00766
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSPRINT
Generic NameDEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Product CodeLWS
Date Received2020-03-31
Model Number6945-65
Catalog Number6945-65
Device Expiration Date2003-02-20
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMPRI
Manufacturer AddressROAD 149 KM 56.3 VILLALBA PR 00766 US 00766


Patients

Patient NumberTreatmentOutcomeDate
15076 2020-03-31

© 2024 FDA.report