RX XIENCE XPEDITION 48 1078300-48

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2020-03-31 for RX XIENCE XPEDITION 48 1078300-48 manufactured by Abbott Vascular.

Event Text Entries

[185725682] The device was not returned for evaluation. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event. Additionally, a review of the complaint history did not indicate a lot specific quality issue. The investigation determined the reported difficulties appear to be related to circumstances of the procedure. There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
Patient Sequence No: 1, Text Type: N, H10


[185725683] It was reported that the procedure was performed to treat a moderately calcified, moderately tortuous right coronary artery that was 90% stenosed. The 3. 0x48mm xience xpedition stent delivery system (sds) failed to cross due to the anatomy and the shaft kinked after repeated attempts were made. It was then noted that the stent was no longer between the balloon markers but still on the balloon. The sds was removed with the stent still on the balloon. The device was replaced with two non-abbott stents (2. 75x24mm and 3. 0x24mm) to successfully continue the procedure. There were no adverse patient effects and no clinically significant delay in the procedure. No additional information was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2024168-2020-03131
MDR Report Key9903326
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-11
Date Mfgr Received2020-03-11
Device Manufacturer Date2019-08-19
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LINDSEY BELL
Manufacturer Street26531 YNEZ RD.
Manufacturer CityTEMECULA CA 925914628
Manufacturer CountryUS
Manufacturer Postal925914628
Manufacturer Phone9519143996
Manufacturer G1ABBOTT VASCULAR, REG # 3005718570 (P099)
Manufacturer StreetCASHEL ROAD
Manufacturer CityCLONMEL TIPPERARY
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRX XIENCE XPEDITION 48
Generic NameDRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Product CodeNIQ
Date Received2020-03-31
Catalog Number1078300-48
Lot Number9081341
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerABBOTT VASCULAR
Manufacturer Address26531 YNEZ RD. TEMECULA CA 925914628 US 925914628


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.