ACTIVA 37601

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2020-03-31 for ACTIVA 37601 manufactured by Medtronic Puerto Rico Operations Co..

Event Text Entries

[185751703] Date of the event (b)(6) 2020 is an estimate. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[185751704] It was reported that about 2 days ago and confirmed "this week," the patient tried to connect to the implantable neurostimulator (ins) and got an out of regulation (oor) on the patient programmer (pp). They were able to bypass the oor once they replaced the batteries in the pp. They have not changed the batteries since they were implanted. Batteries were reviewed. It was also reviewed that heavy duty batteries is not recommended. The patient was redirected to follow up with their healthcare provider (hcp) regarding having the therapy checked. No patient symptoms were reported. Additional information was received from the patient stating that they were still seeing oor on the patient programmer (pp) when patient went to check their settings. It was reviewed for the patient that oor is out of regulation and that the electrical demand on the device is greater than what it was able to deliver. Patient was redirected to the health care provider (hcp) to determine the reason for the oor.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004209178-2020-06451
MDR Report Key9903487
Report SourceCONSUMER
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-16
Date Mfgr Received2020-03-24
Device Manufacturer Date2017-03-07
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLISA WOODWARD CLARK
Manufacturer Street7000 CENTRAL AVENUE NE RCW215
Manufacturer CityMINNEAPOLIS MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7635263920
Manufacturer G1MEDTRONIC PUERTO RICO OPERATIONS CO.
Manufacturer StreetROAD 31, KM. 24, HM 4 CEIBA NORTE INDUSTRIAL PARK
Manufacturer CityJUNCOS PR 00777
Manufacturer CountryUS
Manufacturer Postal Code00777
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameACTIVA
Generic NameSTIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Product CodeMHY
Date Received2020-03-31
Model Number37601
Catalog Number37601
Device Expiration Date2018-08-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC PUERTO RICO OPERATIONS CO.
Manufacturer AddressROAD 31, KM. 24, HM 4 CEIBA NORTE INDUSTRIAL PARK JUNCOS PR 00777 US 00777


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31

© 2026 FDA.report