ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM 24653

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign,health pr report with the FDA on 2020-03-31 for ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM 24653 manufactured by Boston Scientific Corporation.

Event Text Entries

[185944515] This is a combination device. Age at time of event: patient is over 18 years of age. (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[185944516] It was reported that the middle sheath became stuck in the anatomy and the stent partially deployed. A 7x120x130 self expanding drug eluting stent was selected for a procedure in a 75% stenosed lesion in with mild tortuosity the superficial femoral artery (sfa). There was no kink observed via fluoroscopy but the middle sheath became stuck at the lesion, and the physician felt something potentially separated. The thumbwheel was turned to the limit however, the stent only deployed approximately one-third of the way. Another of the same type of device was used to complete this procedure. No patient complications were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-04134
MDR Report Key9903583
Report SourceDISTRIBUTOR,FOREIGN,HEALTH PR
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-13
Date Mfgr Received2020-03-13
Device Manufacturer Date2019-02-22
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAY JOHNSON
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634942574
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer StreetBALLYBRIT BUSINESS PARK
Manufacturer CityGALWAY
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Generic NameSTENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Product CodeNIU
Date Received2020-03-31
Model Number24653
Catalog Number24653
Lot Number0023542000
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer AddressTWO SCIMED PLACE MAPLE GROVE MN 55311 US 55311


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.