AXSYM DRUGS OF ABUSE TOXICOLOGY ASSAY DISK 3D54-08

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2008-02-04 for AXSYM DRUGS OF ABUSE TOXICOLOGY ASSAY DISK 3D54-08 manufactured by Abbott Laboratories.

Event Text Entries

[18883039] This is an initial report. An investigation is in process. A final report will be submitted when the investigation is completed.
Patient Sequence No: 1, Text Type: N, H10


[18969052] A new abbott axsym drugs of abuse (doa) toxicology assay disk, version 8. 0 was released on 11/26/07 which contained a revised axsym. Amphetamine/methamphetamine ii assay file. Installation of the revised assay file is required prior to use of the axsym amphetamine reagents. Complaints have been received indicating when the axsym amphetamine positive and negative interpretation cutoff parameter are edited, error messages are generated before assay runs are completed, resulting in an instrument lock-up. A product correction was issued and reported under 21cfr806 to the fda on february 1, 2008. Abbott has not received any reports of adverse events related to the issue.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1415939-2008-00006
MDR Report Key990366
Report Source07
Date Received2008-02-04
Date of Report2008-01-18
Date of Event2008-01-18
Date Mfgr Received2008-01-18
Date Added to Maude2008-08-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSUDARSHAN HEBBAR,MD, D-3MQC, AP6B
Manufacturer Street100 ABBOTT PARK RD.
Manufacturer CityABBOTT PARK IL 600646112
Manufacturer CountryUS
Manufacturer Postal600646112
Manufacturer Phone8479379677
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction Number1415939-2/01/08-001-C
Event Type3
Type of Report3

Device Details

Brand NameAXSYM DRUGS OF ABUSE TOXICOLOGY ASSAY DISK
Generic NameAMPHETAMINE TEST SYSTEM
Product CodeDKZ
Date Received2008-02-04
Model NumberNA
Catalog Number3D54-08
Lot NumberNA
ID NumberNA
OperatorOTHER
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key992325
ManufacturerABBOTT LABORATORIES
Manufacturer Address100 ABBOTT PARK RD. ABBOTT PARK IL 600643500 US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-02-04

© 2026 FDA.report