PLASMABLADE 3.0S PS210-030S

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-31 for PLASMABLADE 3.0S PS210-030S manufactured by Medtronic Advanced Energy, Llc.

Event Text Entries

[185985175] Product analysis #: 703652291: analysis information 2020-03-12, 11:34:03 cst, pli#: 20; product id#: ps210-030s. Analysis summary: complaint confirmed: upon receiving the device, the heat shrink component was torn and coming off the blade. The device was visually confirmed for this analysis. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[185985176] Information was received from a manufacturing representative (rep) regarding a handpiece. It was reported that the tip coating was coming off the tip of the handpiece. The site opened a new hand piece to continue with the case. There was no patient involvement and the rep was transferred to service and repair.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1226420-2020-00048
MDR Report Key9903774
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-05
Date Mfgr Received2020-03-05
Device Manufacturer Date2019-12-05
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACY RUEMPING
Manufacturer Street7000 CENTRAL AVENUE NE RCW215
Manufacturer CityMINNEAPOLIS MN 55432
Manufacturer CountryUS
Manufacturer Postal55432
Manufacturer Phone7635260594
Manufacturer G1MEDTRONIC ADVANCED ENERGY, LLC
Manufacturer Street180 INTERNATIONAL DRIVE
Manufacturer CityPORTSMOUTH NH 03801
Manufacturer CountryUS
Manufacturer Postal Code03801
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePLASMABLADE 3.0S
Generic NameELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Product CodeGEI
Date Received2020-03-31
Returned To Mfg2020-03-10
Model NumberPS210-030S
Catalog NumberPS210-030S
Lot Number0218983091
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC ADVANCED ENERGY, LLC
Manufacturer Address180 INTERNATIONAL DRIVE PORTSMOUTH NH 03801 US 03801


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31

© 2026 FDA.report