TRIATHLON PRIM CEM FXD BPLT #5 5520-B-500 5520B500

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,other report with the FDA on 2020-03-31 for TRIATHLON PRIM CEM FXD BPLT #5 5520-B-500 5520B500 manufactured by Stryker Orthopaedics-mahwah.

Event Text Entries

[188811828] Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies. There have been no other similar events for the lot referenced. It was noted that the device is not available for evaluation. Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
Patient Sequence No: 1, Text Type: N, H10


[188811829] On (b)(6) 2020, a revision of triathlon that was implanted in 2014. The revision occurred due to aseptic loosening of tibial component. The existing femur component remained stable and was left insitu. The tibia baseplate and insert were removed and replaced with a universal baseplate and 9/100mm stem, tibia cone and cs insert. The procedure doesn? T require monitoring. The outcome of the procedure was a successful revision.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0002249697-2020-00642
MDR Report Key9903865
Report SourceHEALTH PROFESSIONAL,OTHER
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-03
Date Mfgr Received2020-03-03
Device Manufacturer Date2015-01-29
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. COLLIN NEITZEL
Manufacturer Street325 CORPORATE DRIVE
Manufacturer CityMAHWAH NJ 07430
Manufacturer CountryUS
Manufacturer Postal07430
Manufacturer Phone2018315000
Manufacturer G1STRYKER ORTHOPAEDICS-LIMERICK
Manufacturer StreetRAHEEN BUSINESS PARK
Manufacturer CityLIMERICK NA
Manufacturer CountryIE
Manufacturer Postal CodeNA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRIATHLON PRIM CEM FXD BPLT #5
Generic NamePROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Product CodeMBH
Date Received2020-03-31
Model Number5520-B-500
Catalog Number5520B500
Lot NumberEMREJ
Device Expiration Date2020-01-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ORTHOPAEDICS-MAHWAH
Manufacturer Address325 CORPORATE DRIVE MAHWAH NJ 07430 US 07430


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2020-03-31

© 2026 FDA.report