TENDRIL SDX LEAD 1488TC/52

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-31 for TENDRIL SDX LEAD 1488TC/52 manufactured by St. Jude Medical, Inc.(crm-sunnyvale).

Event Text Entries

[185745801] Di not available as product was manufactured on or before september 23, 2014. The results of the investigation are inconclusive since the device was not returned for analysis. Based on the information received, the cause of the reported incident could not be conclusively determined.
Patient Sequence No: 1, Text Type: N, H10


[185745802] Related manufacturer reference number: 2938836-2020-02343, 2938836-2020-02345. It was reported that the patient presented in clinic for device and lead change out procedure. Noise oversensing was observed on the right atrial (ra) lead. The noise was reproduced with any types of patient movements. Due to the patient advanced age and not being candidate for lead explant, the lead was capped on (b)(6) 2020. During the procedure, lead helix damage was observed on the new ra lead. The lead could no longer extend. The lead was cut, and a new sheath was slide over the lead to regain access. The new lead was implanted successfully. The patient was stable.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2938836-2020-02342
MDR Report Key9904192
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-13
Date Mfgr Received2020-03-13
Device Manufacturer Date2006-04-24
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactELIZABETH BOLTZ
Manufacturer Street15900 VALLEY VIEW COURT
Manufacturer CitySYLMAR CA 91342
Manufacturer CountryUS
Manufacturer Postal91342
Manufacturer G1ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
Manufacturer Street645 ALMANOR AVENUE
Manufacturer CitySUNNYVALE CA 94085
Manufacturer CountryUS
Manufacturer Postal Code94085
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTENDRIL SDX LEAD
Generic NamePERMANENT PACEMAKER ELECTRODE
Product CodeNVN
Date Received2020-03-31
Model Number1488TC/52
Lot Number0002145031
Device Expiration Date2009-04-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
Manufacturer Address645 ALMANOR AVENUE SUNNYVALE CA 94085 US 94085


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-31