T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY 1000096 1000886

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2020-03-31 for T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY 1000096 1000886 manufactured by Tandem Diabetes Care.

Event Text Entries

[186853510] The tandem user guide states: do not use any other insulin with your system other than u-100 humalog or novolog. Only humalog and novolog have been tested and found to be compatible for use in the system. No product was returned for evaluation. Should new relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[186853511] It was reported that multiple intermittent occlusion alarms occurred. Reportedly, customer was using fiasp insulin. Tandem technical support educated customer regarding cartridge/insulin labeling. Customer's blood glucose level was 450 mg/dl. Reportedly, the pump supplies were changed to address the issue however, customer reverted to manual injections for insulin therapy.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3013756811-2020-33781
MDR Report Key9904328
Report SourceCONSUMER
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2019-03-13
Date Mfgr Received2020-03-16
Device Manufacturer Date2019-02-01
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMICK TRIER
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8584011451
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameT:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY
Generic NameALTERNATE CONTROLLER ENABLED INFUSION PUMP
Product CodeQFG
Date Received2020-03-31
Model Number1000096
Catalog Number1000886
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTANDEM DIABETES CARE
Manufacturer Address11075 ROSELLE STREET SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31

© 2026 FDA.report