Event Text Entries

[185780288] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[185780289] It was reported that the right ventricular (rv) lead exhibited high and undefined rv coil impedance and high and undefined pacing impedances, both of which have been high since implant. A poor connection of the rv lead to the icd (implantable cardioverter defibrillator) header is suspected. No intervention nor reprogramming has been completed at this time. The rv lead and icd remain in use. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


[186876868] Product event summary: the device was not returned for analysis. However, performance data collected from the device was received and analyzed. Analysis of the device memory indicated the impedance of the right ventricular defibrillation coil was beyond the expected upper range. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9614453-2020-01121
MDR Report Key9904369
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-13
Date Mfgr Received2020-03-14
Device Manufacturer Date2018-12-12
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1MEDTRONIC EUROPE SARL
Manufacturer StreetROUTE DU MOLLIAU 31 CASE POSTALE
Manufacturer CityTOLOCHENAZ VAUD 1131
Manufacturer CountryCH
Manufacturer Postal Code1131
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePROTECTA CRT-D
Generic NameDEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Product CodeNIK
Date Received2020-03-31
Model NumberD364TRG
Catalog NumberD364TRG
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC EUROPE SARL
Manufacturer AddressROUTE DU MOLLIAU 31 CASE POSTALE TOLOCHENAZ VAUD 1131 CH 1131


Patients

Patient NumberTreatmentOutcomeDate
1693565 2020-03-31