OPTISURE ACTIVE FIXATION, DF-4 CONNECTOR LDA210Q/65 LDA210Q-65

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-31 for OPTISURE ACTIVE FIXATION, DF-4 CONNECTOR LDA210Q/65 LDA210Q-65 manufactured by St. Jude Medical, Inc.(crm-sylmar).

Event Text Entries

[185785473] This product is registered as a combination product. The results of the investigation are inconclusive since the device was not returned for analysis. Based on the information received, the cause of the reported incident could not be conclusively determined.
Patient Sequence No: 1, Text Type: N, H10


[185785494] It was reported that the patient presented in clinic. Upon investigation, the right ventricular (rv) lead was noted with intermittent capture, and the rv threshold jumped. Impedance and sensing was stable. The rv lead was explanted and replaced on (b)(6) 2020. The right atrial lead was also electively explanted and replaced during the procedure since the physician didn't want to retain anything in the body after the extraction procedure. The patient was stable.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2017865-2020-04463
MDR Report Key9904514
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-01-22
Date Mfgr Received2020-03-30
Device Manufacturer Date2016-01-19
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactELIZABETH BOLTZ
Manufacturer Street15900 VALLEY VIEW COURT
Manufacturer CitySYLMAR CA 91342
Manufacturer CountryUS
Manufacturer Postal91342
Manufacturer G1ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
Manufacturer Street15900 VALLEY VIEW COURT
Manufacturer CitySYLMAR CA 91342
Manufacturer CountryUS
Manufacturer Postal Code91342
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOPTISURE ACTIVE FIXATION, DF-4 CONNECTOR
Generic NameDEFIBRILLATION LEAD
Product CodeNVY
Date Received2020-03-31
Model NumberLDA210Q/65
Catalog NumberLDA210Q-65
Lot NumberA000016876
Device Expiration Date2017-01-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL, INC.(CRM-SYLMAR)
Manufacturer Address15900 VALLEY VIEW COURT SYLMAR CA 91342 US 91342


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-31

© 2026 FDA.report