Event Text Entries

[185798205] Evaluation summary: post market vigilance (pmv) led an evaluation of one device. An analysis of the system logs noted that towards the end of the logs, the handle was placed on a charger. The next time the handle initializes is with low voltage detected on the charger, a low relative state of charge value, and low individual cell voltages. The handle logs also showed error: handle, hw_error and sd_card_integrity indicating an sd card failure. The handle was opened up and both batteries were found to be vented. A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture. The root cause of the observed condition was determined to be a result of a software fault. The battery cells vented, causing them to leak electrolytic fluid. The battery cells would be unable to charge under these circumstances and would result in the handle entering a cell under voltage state. Improvements have been initiated to mitigate this condition. Additionally, the investigation detected a unreported condition of an sd card integrity error that has no relationship to the reported condition. The system performs sd card integrity check as part of system initialization. This is the only time the check is performed. As a result, the system will fail safely prior to assembly with a clamshell or adapter and poses no patient safety risk. A cross functional team has reviewed the risk and rate of occurrence for this failure mode and has determined that no corrective actions are warranted at this time. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[185798206] According to the reporter, prior the procedure, the handle was being charged in the operation preparation room, but it was found that the red indicator of the charger illuminated. When the handle was removed from the charger to check the situation, liquid scorched part was found at the terminal part of the charger, so charging was stopped after that. When the reported product was collected and checked, although the charger was connected to the outlet, the red indicator flashed. When the handle was connected to a charger that had no problem, it was noted that the indicator flashed green indicating that the handle was being charged, but when the handle was removed from the charger after a while, it could not start up. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1219930-2020-01443
MDR Report Key9904530
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2019-09-13
Date Mfgr Received2020-03-12
Device Manufacturer Date2018-06-12
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLISA HERNANDEZ
Manufacturer Street60 MIDDLETOWN AVE.
Manufacturer CityNORTH HAVEN CT 06473
Manufacturer CountryUS
Manufacturer Postal06473
Manufacturer Phone2034925563
Manufacturer G1COVIDIEN LP LLC NORTH HAVEN
Manufacturer Street195 MCDERMOTT RD
Manufacturer CityNORTH HAVEN CT 06473
Manufacturer CountryUS
Manufacturer Postal Code06473
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSIGNIA
Generic NameSTAPLE, IMPLANTABLE
Product CodeGDW
Date Received2020-03-31
Returned To Mfg2019-10-08
Model NumberSIGPHANDLE
Catalog NumberSIGPHANDLE
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOVIDIEN LP LLC NORTH HAVEN
Manufacturer Address195 MCDERMOTT RD NORTH HAVEN CT 06473 US 06473


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.