DISCOVERY ELBOW 114822

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-31 for DISCOVERY ELBOW 114822 manufactured by Encore Medical L.p..

Event Text Entries

[188208142] The reason for this revision surgery was reported as an infection. The previous surgery and the surgery detailed in this event occurred 1. 3 years apart. The healthcare professional indicated there was no delay in surgery and another suitable device was available for use. The revision surgery was completed as intended. The devices were disposed of at hospital and not made available to djo surgical for examination. A review of the implant device history records (dhr) shows that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements. There were non-conforming material report (ncmr) associated with the main product that may have contributed to the reported event. The devices were within its expiration date at the time of the previous surgery. Non-conforming report history: item: 114822, ncr 12201233, work center 91050, op. Seq. 0165, clean and prep, 1 part scrap. Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review. The root cause of this complaint was a revision surgery due to an infection. There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection. No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system. There are multiple factors that may contribute to an infection that are outside the control of djo surgical. There are no indications of a product or process issue affecting implant safety or effectiveness.
Patient Sequence No: 1, Text Type: N, H10


[188208223] Revision surgery - patient had infection. Removed all original implants. Did not put any of our implants back in.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1644408-2020-00265
MDR Report Key9905080
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-03
Date Mfgr Received2020-03-03
Device Manufacturer Date2018-01-15
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTEFFANY HUTTO
Manufacturer Street9800 METRIC BLVD
Manufacturer CityAUSTIN, TX 78758-5445,
Manufacturer CountryUS
Manufacturer G1ENCORE MEDICAL L.P.
Manufacturer Street9800 METRIC BLVD
Manufacturer CityAUSTIN, TX 78758-5445,
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDISCOVERY ELBOW
Generic NameDISC ULNA 4X75MM LT W BRNG C
Product CodeJDC
Date Received2020-03-31
Catalog Number114822
Lot Number287380
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerENCORE MEDICAL L.P.
Manufacturer Address9800 METRIC BLVD AUSTIN, TX 78758-5445, US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-31

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.