CARESTATION 650 A1

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2020-03-31 for CARESTATION 650 A1 manufactured by Ge Medical Systems (china) Co., Ltd (wuxi).

Event Text Entries

[186174754] A ge healthcare service representative performed a checkout of the system and confirmed the reported issue. Replaced both o2 and n2o sprockets and link chain to resolve the reported issue. No report of patient involvement. The initial reporter is located outside the u. S. , and therefore this information is not provided due to country privacy laws. Unique identifier: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[186174755] The hospital reported a malfunction of the link 25 system causing a hypoxic mix in the patient circuit. There was no report of patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9710602-2020-00117
MDR Report Key9905282
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-05
Date Mfgr Received2020-03-05
Device Manufacturer Date2017-11-27
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactJOHN SZALINSKI
Manufacturer Street3000 N GRANDVIEW BLVD.
Manufacturer CityWAUKESHA WI
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARESTATION 650 A1
Generic NameANESTHESIA GAS MACHINE
Product CodeBSZ
Date Received2020-03-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
Manufacturer AddressNO. 19 CHANGJIANG ROAD NATIONAL HI-TECH DEV. ZONE WUXI 214028 CH 214028


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31