AV RX GRAFTMASTER 1012580-16

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-31 for AV RX GRAFTMASTER 1012580-16 manufactured by Abbott Vascular.

Event Text Entries

[186381427] A visual inspection was performed on the returned device which found the stent to be mislocated. The reported product quality problem/physical property issue was confirmed. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event. Additionally, a review of the complaint history identified no similar incidents from this lot. The investigation determined the reported product quality problem/physical property issue appears to be related to circumstances of the procedure. In this case, it is likely that inadvertent mishandling during unpackaging and/or preparation of the device contributed to the reported proximal and distal stent damage (product quality problem/physical property issue) in addition to the noted stent movement. There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
Patient Sequence No: 1, Text Type: N, H10


[186381428] It was reported that the patient presented with a free perforation with extravasation in the distal left anterior descending artery. Out of packaging, a 2. 80x16mm rx graftmaster was noted to be damaged and therefore not used. The damage was not specified. A 2. 80x16mm rx graftmaster covered stent was deployed at 15 atmospheres and sealed the perforation. There was no reported adverse patient sequela reported. No additional information was provided. Device analysis of the unused 2. 80x16mm rx graftmaster found the stent to be mislocated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2024168-2020-03147
MDR Report Key9905341
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2019-12-09
Date Mfgr Received2020-03-12
Device Manufacturer Date2018-03-14
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LINDSEY BELL
Manufacturer Street26531 YNEZ RD.
Manufacturer CityTEMECULA CA 925914628
Manufacturer CountryUS
Manufacturer Postal925914628
Manufacturer Phone9519143996
Manufacturer G1ABBOTT VASCULAR, REG # 3005718570 (P099)
Manufacturer StreetCASHEL ROAD
Manufacturer CityCLONMEL TIPPERARY
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAV RX GRAFTMASTER
Generic NameCORONARY STENT DELIVERY SYSTEM
Product CodeMAF
Date Received2020-03-31
Returned To Mfg2020-02-24
Model Number1012580-16
Catalog Number1012580-16
Lot Number8031441
Device Expiration Date2020-02-29
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT VASCULAR
Manufacturer Address26531 YNEZ RD. TEMECULA CA 925914628 US 925914628


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31

© 2026 FDA.report