ELLIPSE DR CD2411-36Q

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-31 for ELLIPSE DR CD2411-36Q manufactured by St. Jude Medical, Inc.(crm-sylmar).

Event Text Entries

[185798656] The results of the investigation are inconclusive since the device was not returned for analysis. Based on the information received, the cause of the reported incident could not be conclusively determined.
Patient Sequence No: 1, Text Type: N, H10


[185798657] It was reported that the patient experienced multiple high voltage therapy due to polymorphic ventricular tachycardia and fibrillation. It was noted that the implantable cardioverter defibrillator detected the arrhythmia appropriately but did not sense a few events due to the polymorphic ventricular tachycardia. It was also noted that the device sinus detection was programmed to detect too fast. The rhythm was quickly classified as a sinus rhythm after a high voltage therapy. This caused a delay in therapy because the device had to re-classify the continued arrhythmia. No programming changes were made at this time and the clinic elected to continue to monitor the patient. The patient was reported to be in stable condition.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2017865-2020-04474
MDR Report Key9906303
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-17
Date Mfgr Received2020-03-18
Device Manufacturer Date2018-06-29
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactELIZABETH BOLTZ
Manufacturer Street15900 VALLEY VIEW COURT
Manufacturer CitySYLMAR CA 91342
Manufacturer CountryUS
Manufacturer Postal91342
Manufacturer G1ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
Manufacturer Street15900 VALLEY VIEW COURT
Manufacturer CitySYLMAR CA 91342
Manufacturer CountryUS
Manufacturer Postal Code91342
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameELLIPSE DR
Generic NameIMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Product CodeNVZ
Date Received2020-03-31
Model NumberCD2411-36Q
Catalog NumberCD2411-36Q
Lot NumberA000060776
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL, INC.(CRM-SYLMAR)
Manufacturer Address15900 VALLEY VIEW COURT SYLMAR CA 91342 US 91342


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31

© 2026 FDA.report