MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2020-03-31 for CADD MS3 PUMP 7400 21-7411-51 manufactured by Smiths Medical Asd, Inc..
[185901196]
One smiths medical cadd ms3 pump was returned for analysis in a good condition. During analysis, the customer's reported problem was unable to be duplicated. Inspected the pump and performed functional tests, it passed all tests. Further investigation of the pump found no issue with pump regarding lost programming. Therefore, no problem was found with the issue. Service recommended customer to keep in mind that programming will be lost after 2 days without power from the aaa battery and pay close attention to the low battery notification. Based on the evidence, the complaint was not confirmed, and no fault was found.
Patient Sequence No: 1, Text Type: N, H10
[185901197]
Information was received indicating that a smiths medical cadd ms3 pump lost its programming. No patient was involved in this event. No adverse effects were reported.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3012307300-2020-02484 |
MDR Report Key | 9906427 |
Report Source | DISTRIBUTOR |
Date Received | 2020-03-31 |
Date of Report | 2020-03-31 |
Date Mfgr Received | 2020-03-03 |
Device Manufacturer Date | 2012-05-09 |
Date Added to Maude | 2020-03-31 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | DAVE HALVERSON |
Manufacturer Street | 6000 NATHAN LANE NORTH |
Manufacturer City | MINNEAPOLIS, MN |
Manufacturer Country | US |
Manufacturer G1 | SMITHS MEDICAL ASD, INC. |
Manufacturer Street | 3350 GRANADA AVENUE NORTH |
Manufacturer City | OAKDALE, MN |
Manufacturer Country | US |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CADD MS3 PUMP |
Generic Name | PUMP, INFUSION |
Product Code | FRN |
Date Received | 2020-03-31 |
Returned To Mfg | 2020-03-09 |
Model Number | 7400 |
Catalog Number | 21-7411-51 |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | SMITHS MEDICAL ASD, INC. |
Manufacturer Address | 6000 NATHAN LANE NORTH MINNEAPOLIS, MN US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-31 |