CADD MS3 PUMP 7400 21-7411-51

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2020-03-31 for CADD MS3 PUMP 7400 21-7411-51 manufactured by Smiths Medical Asd, Inc..

Event Text Entries

[185901133] One smiths medical cadd ms3 pump was returned for analysis with a damaged rear housing. During analysis, the customer's reported problem was unable to be duplicated. Inspected the pump and performed functional tests, it passed all test. Further investigation of the pump found no issue with pump regarding lost programming or not holding program. Therefore, no problem found with the issue. Service recommended customer to keep in mind that programming will be lost after 2 days without power from the aaa battery and pay close attention to the low battery notification. Based on the evidence, the complaint was not confirmed, and the problem source of the reported event is unknown.
Patient Sequence No: 1, Text Type: N, H10


[185901134] Information was received indicating that a smiths medical cadd ms3 pump lost its programming and was not holding programming. It was not specified whether this occurred during infusion or interrupted therapy. No adverse effects were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3012307300-2020-02493
MDR Report Key9906431
Report SourceDISTRIBUTOR
Date Received2020-03-31
Date of Report2020-03-31
Date Mfgr Received2020-03-03
Device Manufacturer Date2014-11-10
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAVE HALVERSON
Manufacturer Street6000 NATHAN LANE NORTH
Manufacturer CityMINNEAPOLIS, MN
Manufacturer CountryUS
Manufacturer G1SMITHS MEDICAL ASD, INC.
Manufacturer Street3350 GRANADA AVENUE NORTH
Manufacturer CityOAKDALE, MN
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCADD MS3 PUMP
Generic NamePUMP, INFUSION
Product CodeFRN
Date Received2020-03-31
Returned To Mfg2020-03-10
Model Number7400
Catalog Number21-7411-51
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD, INC.
Manufacturer Address6000 NATHAN LANE NORTH MINNEAPOLIS, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.