SMITHS MEDICAL CADD SOLIS VIP PUMP 2120 21-2120-0102-51

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-03-31 for SMITHS MEDICAL CADD SOLIS VIP PUMP 2120 21-2120-0102-51 manufactured by Smiths Medical Asd, Inc..

Event Text Entries

[185921455] One infusion pump was returned for evaluation. Visual inspection of the device found it to be in good physical condition. A dhr review was performed subsequent to the manufacturing of the device and prior to its release. No problems or issues were identified during this dhr review. Evidence was noted in the event history log: medium alarm displayed: cannot start pump. Device underwent functional testing; the reported customer complaint was confirmed. The faulty air detector was replaced. The problem source has been determined to be unknown.
Patient Sequence No: 1, Text Type: N, H10


[185921456] Information was received that a smiths medical ambulatory infusion cadd solis vip pump had constant air in line error code. Incident occurred during testing; no patient involvment.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3012307300-2020-02402
MDR Report Key9907646
Report SourceUSER FACILITY
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-02-26
Date Mfgr Received2020-03-02
Device Manufacturer Date2013-04-27
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAVE HALVERSON
Manufacturer Street6000 NATHAN LANE N
Manufacturer CityMINNEAPOLIS, MN
Manufacturer CountryUS
Manufacturer G1SMITHS MEDICAL ASD, INC.
Manufacturer Street3350 GRANADA AVENUE NORTH SUITE 100
Manufacturer CityOAKDALE, MN
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSMITHS MEDICAL CADD SOLIS VIP PUMP
Generic NamePUMP, INFUSION
Product CodeFRN
Date Received2020-03-31
Returned To Mfg2020-03-12
Model Number2120
Catalog Number21-2120-0102-51
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD, INC.
Manufacturer Address6000 NATHAN LANE N MINNEAPOLIS, MN US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31