ENDO GIA EGIA60AVM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-31 for ENDO GIA EGIA60AVM manufactured by Covidien Lp Llc North Haven.

Event Text Entries

[186591296] (b)(4). If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[186591297] According to the reporter, when the device was being applied to a closed and detached spleen blood vessels on a laparoscopic splenectomy, the stapler was able to fire however, after firing, the home button could not be returned, the jaws could not be opened, and the device could not be unloaded. The device locked on tissue. Laparoscopic procedure was converted to open surgery, and with another new handle using same reload, a portion of the blood vessel was removed, and the faulty device was cut off with the reload. There was a tissue damaged and enlarged abdominal incision that was extended to more than 1 inch (2. 54cm) as a result of the event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1219930-2020-01460
MDR Report Key9907703
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-03-11
Date Mfgr Received2020-03-13
Device Manufacturer Date2019-04-23
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLISA HERNANDEZ
Manufacturer Street60 MIDDLETOWN AVE.
Manufacturer CityNORTH HAVEN CT 06473
Manufacturer CountryUS
Manufacturer Postal06473
Manufacturer Phone2034925563
Manufacturer G1COVIDIEN LP LLC NORTH HAVEN
Manufacturer Street195 MCDERMOTT RD
Manufacturer CityNORTH HAVEN CT 06473
Manufacturer CountryUS
Manufacturer Postal Code06473
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameENDO GIA
Generic NameSTAPLE, IMPLANTABLE
Product CodeGDW
Date Received2020-03-31
Model NumberEGIA60AVM
Catalog NumberEGIA60AVM
Lot NumberN9D0846KY
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCOVIDIEN LP LLC NORTH HAVEN
Manufacturer Address195 MCDERMOTT RD NORTH HAVEN CT 06473 US 06473


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-31