MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-31 for ENDO GIA EGIA60AVM manufactured by Covidien Lp Llc North Haven.
[186591296]
(b)(4). If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[186591297]
According to the reporter, when the device was being applied to a closed and detached spleen blood vessels on a laparoscopic splenectomy, the stapler was able to fire however, after firing, the home button could not be returned, the jaws could not be opened, and the device could not be unloaded. The device locked on tissue. Laparoscopic procedure was converted to open surgery, and with another new handle using same reload, a portion of the blood vessel was removed, and the faulty device was cut off with the reload. There was a tissue damaged and enlarged abdominal incision that was extended to more than 1 inch (2. 54cm) as a result of the event.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1219930-2020-01460 |
MDR Report Key | 9907703 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-03-31 |
Date of Report | 2020-03-31 |
Date of Event | 2020-03-11 |
Date Mfgr Received | 2020-03-13 |
Device Manufacturer Date | 2019-04-23 |
Date Added to Maude | 2020-03-31 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | LISA HERNANDEZ |
Manufacturer Street | 60 MIDDLETOWN AVE. |
Manufacturer City | NORTH HAVEN CT 06473 |
Manufacturer Country | US |
Manufacturer Postal | 06473 |
Manufacturer Phone | 2034925563 |
Manufacturer G1 | COVIDIEN LP LLC NORTH HAVEN |
Manufacturer Street | 195 MCDERMOTT RD |
Manufacturer City | NORTH HAVEN CT 06473 |
Manufacturer Country | US |
Manufacturer Postal Code | 06473 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ENDO GIA |
Generic Name | STAPLE, IMPLANTABLE |
Product Code | GDW |
Date Received | 2020-03-31 |
Model Number | EGIA60AVM |
Catalog Number | EGIA60AVM |
Lot Number | N9D0846KY |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | * |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | COVIDIEN LP LLC NORTH HAVEN |
Manufacturer Address | 195 MCDERMOTT RD NORTH HAVEN CT 06473 US 06473 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Required No Informationntervention | 2020-03-31 |