PUMP MMT-1715KM 630G 3ML BLACK MEDI ?

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2020-03-31 for PUMP MMT-1715KM 630G 3ML BLACK MEDI ? manufactured by Medtronic Puerto Rico Operations Co..

Event Text Entries

[186201685] Pump passed sleep current measurement, active current measurement, self test and displacement test. However, pump trace download analysis confirmed the pump alarmed power error 25, power loss alarm and replace battery alert/replace battery now alarm. The power management tool confirmed power error 25, power loss alarm and replace battery alert/replace battery now alarm was triggered when the backup battery loaded voltage (loaded vlith) was less than 3. 5v for 4 consecutive hours due to connector resistance j6 /pcb 1. After disconnecting and reconnecting the internal battery connector on j6/pcb 1, the pump was monitored and functioned properly.
Patient Sequence No: 1, Text Type: N, H10


[186201686] Information received by medtronic indicated that the insulin pump had replace battery alarm and also they received low battery alert. Troubleshooting was performed. No harm requiring medical intervention was reported. The insulin pump will be returned for analysis.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004209178-2020-92373
MDR Report Key9907754
Report SourceCONSUMER
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-01-07
Date Mfgr Received2020-03-20
Device Manufacturer Date2017-01-10
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactGERWIN DE GRAAFF
Manufacturer StreetCEIBA NORTE IND. PARK #50 ROAD
Manufacturer CityJUNCOS 00777-386
Manufacturer Country*
Manufacturer Postal00777-3869
Manufacturer Phone8185464805
Manufacturer G1MEDTRONIC PUERTO RICO OPERATIONS CO.
Manufacturer StreetCEIBA NORTE IND. PARK #50 ROAD
Manufacturer CityJUNCOS 00777-386
Manufacturer Country*
Manufacturer Postal Code00777-3869
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePUMP MMT-1715KM 630G 3ML BLACK MEDI ?
Generic NameARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Product CodeOZO
Date Received2020-03-31
Returned To Mfg2020-01-27
Model NumberMMT-1715KM
Catalog NumberMMT-1715KM
Lot NumberHG1L3EP
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC PUERTO RICO OPERATIONS CO.
Manufacturer AddressCEIBA NORTE IND. PARK #50 ROAD JUNCOS 00777-386 * 00777-3869


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31