GLOBAL UNITE HEAD 48X18 ECC 1100-48-610 110048610

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-31 for GLOBAL UNITE HEAD 48X18 ECC 1100-48-610 110048610 manufactured by Depuy Ireland - 9616671.

Event Text Entries

[186797942] (b)(4). If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
Patient Sequence No: 1, Text Type: N, H10


[186797943] Patient was presented with pain in abduction. Dr. Carried out a scan of the effected shoulder and printed a 3d model of the shoulder, he determined that the humeral head eccentricity was incorrectly placed resulting in the greater tuberosity articulating against the subacromial in abduction. Revision surgery was preformed with humeral head removed and replaced with the same size ecc 48 x 18mm and the eccentricity changed from inferior medial position to superior lateral position. Dr. Carried out range of motion test and determined that greater tuberosity no longer articulated against the subacromial.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1818910-2020-09799
MDR Report Key9907848
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-31
Date of Report2020-03-17
Date of Event2020-03-17
Date Mfgr Received2020-03-17
Device Manufacturer Date2017-01-20
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKARA DITTY-BOVARD
Manufacturer Street700 ORTHOPAEDIC DRIVE
Manufacturer CityWARSAW IN 465810988
Manufacturer CountryUS
Manufacturer Postal465810988
Manufacturer Phone6107428552
Manufacturer G1DEPUY ORTHOPAEDICS, INC. 1818910
Manufacturer Street700 ORTHOPAEDIC DRIVE
Manufacturer CityWARSAW IN 465810988
Manufacturer CountryUS
Manufacturer Postal Code465810988
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGLOBAL UNITE HEAD 48X18 ECC
Generic NameGLOBAL UNITE IMPLANTS : HUMERAL TRIALS
Product CodeHSD
Date Received2020-03-31
Model Number1100-48-610
Catalog Number110048610
Lot NumberC32541
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDEPUY IRELAND - 9616671
Manufacturer AddressLOUGHBEG RINGASKIDDY CO. CORK EI


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-31

© 2026 FDA.report