MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-31 for S7 STEALTHSTATION NAVIGATION SYSTEM 9733856 manufactured by Medtronic Navigation, Inc.
[185951796]
A software analysis was initiated. Analysis found that the reported behavior was the intended functionality of the software. The software functioned as designed. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[185951797]
Medtronic received information that, while outside of a procedure, the site was having issues with seeing surgeon profiles and procedures in the spine application software. It was reported that the site checked administrative settings and an error message was received. There was no patient present when this issue was identified. It was late reported that the imaging system tool card had been removed from the navigation system, which subsequently removed all procedures and profiles. To restore functionality, the tool card was re-added and tools were re-installed. Medtronic received information that the issue occurred during testing.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1723170-2020-01119 |
MDR Report Key | 9908000 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-03-31 |
Date of Report | 2020-03-31 |
Date of Event | 2020-03-19 |
Date Mfgr Received | 2020-03-19 |
Date Added to Maude | 2020-03-31 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | BIOMEDICAL ENGINEER |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | STACY RUEMPING |
Manufacturer Street | 7000 CENTRAL AVENUE NE RCW215 |
Manufacturer City | MINNEAPOLIS MN 55432 |
Manufacturer Country | US |
Manufacturer Postal | 55432 |
Manufacturer Phone | 7635260594 |
Manufacturer G1 | MEDTRONIC NAVIGATION, INC |
Manufacturer Street | 826 COAL CREEK CIRCLE |
Manufacturer City | LOUISVILLE CO 80027 |
Manufacturer Country | US |
Manufacturer Postal Code | 80027 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | S7 STEALTHSTATION NAVIGATION SYSTEM |
Generic Name | INSTRUMENT, STEREOTAXIC |
Product Code | HAW |
Date Received | 2020-03-31 |
Model Number | 9733856 |
Catalog Number | 9733856 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MEDTRONIC NAVIGATION, INC |
Manufacturer Address | 826 COAL CREEK CIRCLE LOUISVILLE CO 80027 US 80027 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-31 |