MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-31 for EEA EEA28 manufactured by Us Surgical Puerto Rico.
[186106797]
(b)(4). If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[186106798]
According to the reporter, following a rectal anastomoses, after the gun was fired on a laparoscopic sigmoid colectomy procedure, the stapler was stuck and was difficult to be removed from the cavity. The stapler was removed by hand manually but the anvil was left inside and a formation of ileostomy was carried out to prevent any further complications. The head of the anvil was pushed from the top and was grabbed with a babcock and pulled out. The two doughnuts were okay but the rectal one was thin but intact. The patient was discharged with no problem and will have watersolube enema to check the anastomosis in 2-3 weeks.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2647580-2020-01155 |
MDR Report Key | 9908197 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-03-31 |
Date of Report | 2020-03-31 |
Date of Event | 2020-03-06 |
Date Mfgr Received | 2020-03-12 |
Device Manufacturer Date | 2019-03-25 |
Date Added to Maude | 2020-03-31 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | LISA HERNANDEZ |
Manufacturer Street | 60 MIDDLETOWN AVE |
Manufacturer City | NORTH HAVEN CT 06473 |
Manufacturer Country | US |
Manufacturer Postal | 06473 |
Manufacturer Phone | 2034925563 |
Manufacturer G1 | US SURGICAL PUERTO RICO |
Manufacturer Street | 201 SABANETAS INDUSTRIAL PARK |
Manufacturer City | PONCE PR 007164401 |
Manufacturer Country | US |
Manufacturer Postal Code | 007164401 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | EEA |
Generic Name | STAPLE, IMPLANTABLE |
Product Code | GDW |
Date Received | 2020-03-31 |
Returned To Mfg | 2020-03-20 |
Model Number | EEA28 |
Catalog Number | EEA28 |
Lot Number | P9C1603KY |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | US SURGICAL PUERTO RICO |
Manufacturer Address | 201 SABANETAS INDUSTRIAL PARK PONCE PR 007164401 US 007164401 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Required No Informationntervention | 2020-03-31 |