MAUDE MDR 9908231

MDR report key
9908231
Report number
2024168-2020-03171
Event key
0
Event type
3
Date of event
2020-02-01
Date received
2020-03-31
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
1
Health professional
3
Initial report to FDA
3
Event location
3

Manufacturer Contact#

Contact
MS. LINDSEY BELL
Address
26531 YNEZ RD. TEMECULA CA 92591 US
Phone
951-951-9519
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1STARCLOSE VASCULARY CLOSURE SYSTEMVESSEL CLOSURE CLIPABBOTT VASCULARMGB14679-0114679-01R N

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12020-03-3101. R

Event Narratives#

N

Patient 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. A REVIEW OF THE LOT HISTORY RECORD AND COMPLAINT HISTORY OF THE REPORTED LOT COULD NOT BE CONDUCTED BECAUSE THE LOT NUMBER WAS NOT PROVIDED. THE INVESTIGATION WAS UNABLE TO DETERMINE A CONCLUSIVE CAUSE FOR THE REPORTED DIFFICULTIES; HOWEVER, THE TREATMENT APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING.

D

Patient 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF A COMMON FEMORAL ARTERY WAS ATTEMPTED USING A STARCLOSE SE DEVICE WITH A 6F SHEATH AFTER AN INTERVENTIONAL LOW LEG PROCEDURE. REPORTEDLY, THE STARCLOSE SE DEVICE DID NOT ACHIEVE HEMOSTASIS. MANUAL ARTERIAL COMPRESSION WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.