DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM 9500-45 STS-OM-003

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2020-03-31 for DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM 9500-45 STS-OM-003 manufactured by Dexcom, Inc..

Event Text Entries

[186550145] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[186550146] It was reported that a detached sensor wire occurred. The sensor was inserted at the abdomen on (b)(6) 2020. It was indicated that the patient removed the transmitter from the sensor pod before removing the sensor from the body, which is misuse of the device. No product was provided for evaluation. Confirmation of the allegation and a probable cause could not be determined. No injury or medical intervention was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004753838-2020-031095
MDR Report Key9908308
Report SourceCONSUMER
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-02-28
Date Mfgr Received2020-03-09
Device Manufacturer Date2019-10-20
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS ASHLEY SPOTO
Manufacturer Street6340 SEQUENCE DR.
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8582000200
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM
Generic NameCONTINUOUS GLUCOSE MONITOR
Product CodeQBJ
Date Received2020-03-31
Model Number9500-45
Catalog NumberSTS-OM-003
Lot Number5263768
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDEXCOM, INC.
Manufacturer Address6340 SEQUENCE DR. SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31

© 2020 FDA.report