MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2020-03-31 for MEDFUSION PUMP 3500 3500-500 manufactured by Smiths Medical Asd, Inc..
[185904423]
One medfusion pump was received in good physical condition with no appearance of any physical damage. The event history log was reviewed and there was evidence of a primary audible alarm background test. The power up process was conducted and the reported issue was unable to be replicated. No physical or any other damage was found on the speaker or interconnect board. The speaker was replaced as a preventative measure. The force sensor, interconnect board and battery were also replaced.
Patient Sequence No: 1, Text Type: N, H10
[185904424]
Information was received indicating that a smiths medical medfusion pump had a primary alarm failure. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3012307300-2020-02629 |
MDR Report Key | 9908369 |
Report Source | CONSUMER |
Date Received | 2020-03-31 |
Date of Report | 2020-03-31 |
Date Mfgr Received | 2020-03-06 |
Device Manufacturer Date | 2011-05-24 |
Date Added to Maude | 2020-03-31 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | DAVE HALVERSON |
Manufacturer Street | 6000 NATHAN LANE NORTH |
Manufacturer City | MINNEAPOLIS, MN |
Manufacturer Country | US |
Manufacturer Phone | 3833310 |
Manufacturer G1 | SMITHS MEDICAL ASD, INC. |
Manufacturer Street | 3350 GRANADA AVENUE NORTH SUITE 100 |
Manufacturer City | OAKDALE, MN |
Manufacturer Country | US |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | MEDFUSION PUMP |
Generic Name | PUMP, INFUSION |
Product Code | FRN |
Date Received | 2020-03-31 |
Returned To Mfg | 2020-03-13 |
Model Number | 3500 |
Catalog Number | 3500-500 |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | SMITHS MEDICAL ASD, INC. |
Manufacturer Address | 6000 NATHAN LANE NORTH MINNEAPOLIS, MN US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-31 |