N
Patient 1
(B)(4). UNIT PASSED DISPLACEMENT TEST, SELF TEST AND ACTIVE CURRENT MEASUREMENT. HOWEVER, UNIT FAILED SLEEP CURRENT MEASUREMENT AND POWER MANAGEMENT GRAPH DISPLAYS UNEXPECTED BATTERY POWER LOSS, PROBLEM ISOLATED TO ELECTRONIC ASSEMBLIES.
| Seq | Brand | Generic | Manufacturer | Product code | Model | Catalog | Lot | PMA | 510(k) | Implant | Evaluated | Availability |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-1781K 670G V4.11 MG | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | MEDTRONIC PUERTO RICO OPERATIONS CO. | OZP | MMT-1781K | MMT-1781K | HG316XA | R | R |
| Sequence | Received | Treatment | Outcome |
|---|---|---|---|
| 1 | 2020-03-31 | 0 |
Patient 1
(B)(4). UNIT PASSED DISPLACEMENT TEST, SELF TEST AND ACTIVE CURRENT MEASUREMENT. HOWEVER, UNIT FAILED SLEEP CURRENT MEASUREMENT AND POWER MANAGEMENT GRAPH DISPLAYS UNEXPECTED BATTERY POWER LOSS, PROBLEM ISOLATED TO ELECTRONIC ASSEMBLIES.
Patient 1
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP BATTERY KEEPS LASTING ONLY 2-3 DAYS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.