PUMP MMT-1781K 670G V4.11 MG

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-31 for PUMP MMT-1781K 670G V4.11 MG manufactured by Medtronic Puerto Rico Operations Co..

Event Text Entries

[186162979] (b)(4). Unit passed displacement test, self test and active current measurement. However, unit failed sleep current measurement and power management graph displays unexpected battery power loss, problem isolated to electronic assemblies.
Patient Sequence No: 1, Text Type: N, H10


[186162980] Information received by medtronic indicated that the insulin pump battery keeps lasting only 2-3 days. No harm requiring medical intervention was reported. The insulin pump will be returned for analysis.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004209178-2020-92473
MDR Report Key9908884
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2019-08-20
Date Mfgr Received2020-03-23
Device Manufacturer Date2018-12-12
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactGERWIN DE GRAAFF
Manufacturer StreetCEIBA NORTE IND. PARK #50 ROAD
Manufacturer CityJUNCOS 00777-386
Manufacturer Country*
Manufacturer Postal00777-3869
Manufacturer Phone8185464805
Manufacturer G1MEDTRONIC PUERTO RICO OPERATIONS CO.
Manufacturer StreetCEIBA NORTE IND. PARK #50 ROAD
Manufacturer CityJUNCOS 00777-386
Manufacturer Country*
Manufacturer Postal Code00777-3869
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePUMP MMT-1781K 670G V4.11 MG
Generic NameARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Product CodeOZP
Date Received2020-03-31
Returned To Mfg2020-01-31
Model NumberMMT-1781K
Catalog NumberMMT-1781K
Lot NumberHG316XA
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC PUERTO RICO OPERATIONS CO.
Manufacturer AddressCEIBA NORTE IND. PARK #50 ROAD JUNCOS 00777-386 * 00777-3869


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31

© 2024 FDA.report