MAUDE MDR 9908884

MDR report key
9908884
Report number
3004209178-2020-92473
Event key
0
Event type
3
Date of event
2019-08-20
Date received
2020-03-31
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
0
Health professional
3
Initial report to FDA
3
Event location
3

Manufacturer Contact#

Contact
GERWIN DE GRAAFF
Address
CEIBA NORTE IND. PARK #50 ROAD JUNCOS 00777 *
Phone
818-818-8185
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1PUMP MMT-1781K 670G V4.11 MGARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROLMEDTRONIC PUERTO RICO OPERATIONS CO.OZPMMT-1781KMMT-1781KHG316XAR R

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12020-03-310

Event Narratives#

N

Patient 1

(B)(4). UNIT PASSED DISPLACEMENT TEST, SELF TEST AND ACTIVE CURRENT MEASUREMENT. HOWEVER, UNIT FAILED SLEEP CURRENT MEASUREMENT AND POWER MANAGEMENT GRAPH DISPLAYS UNEXPECTED BATTERY POWER LOSS, PROBLEM ISOLATED TO ELECTRONIC ASSEMBLIES.

D

Patient 1

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP BATTERY KEEPS LASTING ONLY 2-3 DAYS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.