670G INSULIN PUMP MMT-1780KL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2020-03-31 for 670G INSULIN PUMP MMT-1780KL manufactured by Medtronic Puerto Rico Operations Co..

Event Text Entries

[186372583] Unit received with missing retainer, missing reservoir tube o-ring and broken reservoir tube lip. Unable to perform displacement test or lock reservoir in place due to missing retainer. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[186372584] Information received by medtronic indicated that the retainer ring was cracked and broken. The customer stated that the reservoir was able to lock into place. No harm requiring medical intervention was reported. The insulin pump will be returned for analysis.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004209178-2020-92478
MDR Report Key9908902
Report SourceCONSUMER
Date Received2020-03-31
Date of Report2020-03-31
Date of Event2020-01-27
Date Mfgr Received2020-03-23
Device Manufacturer Date2019-04-16
Date Added to Maude2020-03-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactGERWIN DE GRAAFF
Manufacturer StreetCEIBA NORTE IND. PARK #50 ROAD
Manufacturer CityJUNCOS 00777-386
Manufacturer Country*
Manufacturer Postal00777-3869
Manufacturer Phone8185464805
Manufacturer G1MEDTRONIC PUERTO RICO OPERATIONS CO.
Manufacturer StreetCEIBA NORTE IND. PARK #50 ROAD
Manufacturer CityJUNCOS 00777-386
Manufacturer Country*
Manufacturer Postal Code00777-3869
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name670G INSULIN PUMP MMT-1780KL
Generic NameARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Product CodeOZP
Date Received2020-03-31
Returned To Mfg2020-02-05
Model NumberMMT-1780KL
Catalog NumberMMT-1780KL
Lot NumberHG3BYFK
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC PUERTO RICO OPERATIONS CO.
Manufacturer AddressCEIBA NORTE IND. PARK #50 ROAD JUNCOS 00777-386 * 00777-3869


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-31

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.