MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-31 for 640G INSULIN PUMP MMT-1712K manufactured by Medtronic Puerto Rico Operations Co..
[186372141]
(b)(4). Unit p-cap / reservoir locked properly in place. Unit received with cracked lcd window. Unit received with critical pump error due to moisture damage to force sensor. Unit received with corroded electronic assemblies, corroded battery tube, and corroded motor home switch. Unit unable to perform displacement test, rewind, prime/seating, basic occlusion, force sensor, occlusion, sleep current measurement, active current measurement, and self test due to critical pump error (open book image) alarm. The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states. However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
Patient Sequence No: 1, Text Type: N, H10
[186372142]
Information received by medtronic indicated that the insulin pump? S crack was seen on display. Customer stated that reservoir able to lock in place when insert into pump. No harm requiring medical intervention was reported. The insulin pump will be returned for analysis.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3004209178-2020-92489 |
MDR Report Key | 9908925 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-03-31 |
Date of Report | 2020-03-31 |
Date of Event | 2019-10-29 |
Date Mfgr Received | 2020-03-23 |
Device Manufacturer Date | 2018-03-09 |
Date Added to Maude | 2020-03-31 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | GERWIN DE GRAAFF |
Manufacturer Street | CEIBA NORTE IND. PARK #50 ROAD |
Manufacturer City | JUNCOS 00777-386 |
Manufacturer Country | * |
Manufacturer Postal | 00777-3869 |
Manufacturer Phone | 8185464805 |
Manufacturer G1 | MEDTRONIC PUERTO RICO OPERATIONS CO. |
Manufacturer Street | CEIBA NORTE IND. PARK #50 ROAD |
Manufacturer City | JUNCOS 00777-386 |
Manufacturer Country | * |
Manufacturer Postal Code | 00777-3869 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | 640G INSULIN PUMP MMT-1712K |
Generic Name | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR |
Product Code | OYC |
Date Received | 2020-03-31 |
Returned To Mfg | 2020-02-05 |
Model Number | MMT-1712K |
Catalog Number | MMT-1712K |
Lot Number | HG2DG7M |
Operator | LAY USER/PATIENT |
Device Availability | R |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MEDTRONIC PUERTO RICO OPERATIONS CO. |
Manufacturer Address | CEIBA NORTE IND. PARK #50 ROAD JUNCOS 00777-386 * 00777-3869 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-31 |