MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-31 for 640G INSULIN PUMP MMT-1712K manufactured by Medtronic Puerto Rico Operations Co..
[185908809]
(b)(4). Unit received with missing retainer ring and reservoir tube lip o-ring. Unit received with partially broken off piece at the reservoir tube lip. Unable to perform displacement test and p-cap / reservoir will not lock properly due to missing retainer. Unit received with cracked case (battery tube), cracked case-corner of belt clip rails, and cracked battery tube threads. The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states. However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
Patient Sequence No: 1, Text Type: N, H10
[185908810]
Information received by medtronic indicated that the insulin pump had a detached retainer ring. The customer stated that the reservoir did lock into place. Troubleshooting was performed for damaged. No harm requiring medical intervention was reported. The insulin pump will be returned for analysis.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3004209178-2020-92501 |
MDR Report Key | 9908969 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-03-31 |
Date of Report | 2020-03-31 |
Date of Event | 2019-12-20 |
Date Mfgr Received | 2020-03-23 |
Device Manufacturer Date | 2017-10-30 |
Date Added to Maude | 2020-03-31 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | GERWIN DE GRAAFF |
Manufacturer Street | CEIBA NORTE IND. PARK #50 ROAD |
Manufacturer City | JUNCOS 00777-386 |
Manufacturer Country | * |
Manufacturer Postal | 00777-3869 |
Manufacturer Phone | 8185464805 |
Manufacturer G1 | MEDTRONIC PUERTO RICO OPERATIONS CO. |
Manufacturer Street | CEIBA NORTE IND. PARK #50 ROAD |
Manufacturer City | JUNCOS 00777-386 |
Manufacturer Country | * |
Manufacturer Postal Code | 00777-3869 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | 640G INSULIN PUMP MMT-1712K |
Generic Name | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR |
Product Code | OYC |
Date Received | 2020-03-31 |
Returned To Mfg | 2020-02-05 |
Model Number | MMT-1712K |
Catalog Number | MMT-1712K |
Operator | LAY USER/PATIENT |
Device Availability | R |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | MEDTRONIC PUERTO RICO OPERATIONS CO. |
Manufacturer Address | CEIBA NORTE IND. PARK #50 ROAD JUNCOS 00777-386 * 00777-3869 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-31 |