AXSYM DRUGS OF ABUSE TOXICOLOGY ASSAY DISK 3D54-08

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2008-02-05 for AXSYM DRUGS OF ABUSE TOXICOLOGY ASSAY DISK 3D54-08 manufactured by Abbott Laboratories.

Event Text Entries

[802987] The customer stated vrtx error 002 in task 40 occurred on the axsym analyzer three times after installing the drugs of abuse (doa) assay disk version 8. 0 and calibrating the assay. The customer cycled power and reinstalled the software without issue, but encountered the error again when running the amphetamine assay. Abbot field service formatted the hard drive, reloaded axsym system software version 6. 0, and reloaded the doa assay disk version 8. 0. When the cutoff parameters were adjusted after service, the error reoccurred. A copy of the system software in use was returned to abbott for further investigation. No impact to pt management was reported.
Patient Sequence No: 1, Text Type: D, B5


[8045292] This is an initial report. An investigation is in process. A final report will be submitted when the investigation is completed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1415939-2008-00010
MDR Report Key991046
Report Source04
Date Received2008-02-05
Date of Report2008-01-15
Date of Event2007-12-20
Date Mfgr Received2008-01-15
Date Added to Maude2008-07-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactSUDARSHAN HEBBAR, MD, D-3MQC, AP6B
Manufacturer Street100 ABBOTT PARK RD.
Manufacturer CityABBOTT PARK IL 600646112
Manufacturer CountryUS
Manufacturer Postal600646112
Manufacturer Phone8479379677
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction Number1415939-2/01/08-001-C
Event Type3
Type of Report3

Device Details

Brand NameAXSYM DRUGS OF ABUSE TOXICOLOGY ASSAY DISK
Generic NameAMPHETAMINE TEST SYSTEM
Product CodeMGX
Date Received2008-02-05
Model NumberNA
Catalog Number3D54-08
Lot NumberNA
ID NumberVERSION 8.0
OperatorOTHER
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key992323
ManufacturerABBOTT LABORATORIES
Manufacturer Address100 ABBOTT PARK RD. ABBOTT PARK IL 600643500 US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-02-05

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.