BBL THIOGLYCOLLATE MEDIUM WITH CALCIUM CARBO 298518

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2008-02-01 for BBL THIOGLYCOLLATE MEDIUM WITH CALCIUM CARBO 298518 manufactured by Bd Microbiology Systems.

Event Text Entries

[804345] A technician was attempting to unscrew the cap of a tube to inoculate the media with a specimen. The cap was tight and the technician broke the tube while trying to open it. The breakage resulted in a cut on the technician's hand which required stitches.
Patient Sequence No: 1, Text Type: D, B5


[7946669] No returns were received for evaluation. Retention material from this lot of material was examined for tight caps. Examination with a thread kit found all material acceptable with no defects to the glass. Torque testing confirmed torque value above the initial release criteria. Manufacturing and release testing records were examined and found to be acceptable. Scheduled preventive maintenance is done on the torque machine and was current during manufacture of this lot of material. Root cause is suspected to be operator error. As corrective action, the complaint report was reviewed with the operators responsible for manufacturing this material. Retraining on proper torquing procedures occurred and has been documented. We will continue to monitor this issue for trends.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1119779-2008-00002
MDR Report Key991293
Report Source05
Date Received2008-02-01
Date of Report2008-02-01
Date of Event2008-01-04
Date Mfgr Received2008-01-04
Device Manufacturer Date2007-11-01
Date Added to Maude2008-02-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJOHN GERLICH
Manufacturer Street7 LOVETON CIRCLE
Manufacturer CitySPARKS MD 21152
Manufacturer CountryUS
Manufacturer Postal21152
Manufacturer Phone4103164464
Manufacturer G1BD MICROBIOLOGY SYSTEMS
Manufacturer Street2801 INDUSTRIAL DR.
Manufacturer CityMADISON WI 53713
Manufacturer CountryUS
Manufacturer Postal Code53713
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBBL THIOGLYCOLLATE MEDIUM WITH CALCIUM CARBO
Product CodeJSG
Date Received2008-02-01
Catalog Number298518
Lot Number7309612
Device Expiration Date2008-11-07
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key960696
ManufacturerBD MICROBIOLOGY SYSTEMS
Manufacturer Address2801 INDUSTRIAL DR. MADISON WI 53713 US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2008-02-01

© 2026 FDA.report