GENESEARCH BREAST LYMPH NODE ASSAY 2900005

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 99 report with the FDA on 2008-02-14 for GENESEARCH BREAST LYMPH NODE ASSAY 2900005 manufactured by Veridex Llc.

Event Text Entries

[780184] Customer complaint about sprays of ethanol coming out of vacuum pump. Issue is sporadic. Customer empties the manifold after each run and nevertheless experience the issue. Customer says spray is blue colored and sometimes hot. Customer did not experience any injuries or burns. No medical treatment was required.
Patient Sequence No: 1, Text Type: D, B5


[8092377] Vacuum pump and qiagan qiavac-24 plus are not a device (general lab equipment). Device did not malfunction. User set-up of vacuum and pump may not be correct. No medical injury or treatment. Labeling for device (warning and precaution #16) states: all personnel should follow universal precautions and use lab safety equipment (i. E. Safety glasses, lab coat, gloves). If available, a fume hood should be used for preparing sample preparation kit reagents, and a biosafety cabinet should be used for preparing/handling biological material if a hood is not available, consider using a face shield during the homogenization step.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004582358-2008-00001
MDR Report Key998834
Report Source99
Date Received2008-02-14
Date of Report2008-02-14
Date of Event2008-01-18
Date Mfgr Received2008-01-18
Date Added to Maude2009-04-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDEBRA RASMUSSEN
Manufacturer Street1001 US HWY 202
Manufacturer CityRARITAN NJ 08869
Manufacturer CountryUS
Manufacturer Postal08869
Manufacturer Phone9085415843
Manufacturer G1VERIDEX, LLC
Manufacturer Street33 TECHNOLOGY DR.
Manufacturer CityWARREN NJ 07059
Manufacturer CountryUS
Manufacturer Postal Code07059
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGENESEARCH BREAST LYMPH NODE ASSAY
Generic NameGENESEARCH SAMPLE PREPARATION KIT
Product CodeOCB
Date Received2008-02-14
Model NumberNA
Catalog Number2900005
Lot NumberNOT PROVIDED
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerVERIDEX LLC
Manufacturer AddressWARREN NJ US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-02-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.