PMA P820076S020

Device
BIOTRONIK GEMNOS 04, 04A, AND TC 04
Applicant
Biotronik, Inc.
PMA number
P820076
Supplement
S020
Product code
LWP
Decision date
1995-12-22
Classification
Implantable Pulse Generator, Pacemaker (non-crt)
Generic name
Implantable pulse generator, pacemaker (non-CRT)
Approval order statement
APPROVAL FOR THE CHANGE IN THE BURN-IN PROCESSING OF THE WILSON GREATBATCH BATTERIES WHICH ARE USED IN YOUR LISTED PACEMAKERS. THIS PROCESS CHANGE IS A REDUCTION OF THE BATTERY BURN-IN TIME FROM 35 DAYS TO 4 DAYS

Current openFDA PMA Record#

Device
BIOTRONIK GEMNOS 04, 04A, AND TC 04
Applicant
Biotronik, Inc.
PMA number
P820076
Supplement
S020
Product code
LWP
Generic name
Implantable pulse generator, pacemaker (non-CRT)
Decision date
1995-12-22
Decision code
APPR
Date received
1995-05-04
Supplement type
Normal 180 Day Track
Supplement reason
Change Design/Components/Specifications/Material
Approval order statement
APPROVAL FOR THE CHANGE IN THE BURN-IN PROCESSING OF THE WILSON GREATBATCH BATTERIES WHICH ARE USED IN YOUR LISTED PACEMAKERS. THIS PROCESS CHANGE IS A REDUCTION OF THE BATTERY BURN-IN TIME FROM 35 DAYS TO 4 DAYS