PMA P860004S167

Device
INDURA 1 PIECE INTRATHECAL CATHETER WITH SUTURELESS CONNECTOR/INTRATHECAL CATHETER WITH SUTURELESS CONNECTOR
Applicant
Medtronic, Inc.
PMA number
P860004
Supplement
S167
Product code
LKK
Decision date
2012-02-17
Classification
Pump, Infusion, Implanted, Programmable
Generic name
Pump, infusion, implanted, programmable
Approval order statement
APPROVAL FOR THE SUTURELESS CONNECTOR (SC) CATHETER MODELS 8709SC AND 8731SC, AND REVISION KIT MODELS 8578 AND 8596SC TO ADDRESS THE POTENTIAL FOR SIDE LOAD LEAKS. THE MODIFICATION TO YOUR SUTURELESS CONNECTOR ARE AS FOLLOWS: 1) DECREASED THE COUNTER BORE DIAMETER FROM 0.040" TO 0.035"; 2) DECREASED THE DEPTH OF THE COUNTER BORE FROM 0.055" TO 0.050"; 3) INCREASED THE TOLERANCE FOR THE 0.15" DIMENSION OF THE TUBING POSITION FROM +-0.01 TO +-0.02; AND 4) ADDED A REQUIREMENT TO THE PRODUCT SPECIFICATION FOR SIDE LOAD FORCES: CONNECTOR MUST NOT OCCLUDE OR LEAK WHEN PRESSURIZED TO 1.0 +- 0.1 PSIG AIR PRESSURE AND A FORCE OF 4.8 +- 0.1 LBF IS DISTRIBUTED OVER THE CONNECTOR SURFACE AT AN ANGLE OF 12 DEGREES +- 2 DEGREES.

Current openFDA PMA Record#

Device
INDURA 1 PIECE INTRATHECAL CATHETER WITH SUTURELESS CONNECTOR/INTRATHECAL CATHETER WITH SUTURELESS CONNECTOR
Applicant
Medtronic, Inc.
PMA number
P860004
Supplement
S167
Product code
LKK
Generic name
Pump, infusion, implanted, programmable
Decision date
2012-02-17
Decision code
APPR
Date received
2011-12-27
Supplement type
Real-Time Process
Supplement reason
Change Design/Components/Specifications/Material
Approval order statement
APPROVAL FOR THE SUTURELESS CONNECTOR (SC) CATHETER MODELS 8709SC AND 8731SC, AND REVISION KIT MODELS 8578 AND 8596SC TO ADDRESS THE POTENTIAL FOR SIDE LOAD LEAKS. THE MODIFICATION TO YOUR SUTURELESS CONNECTOR ARE AS FOLLOWS: 1) DECREASED THE COUNTER BORE DIAMETER FROM 0.040" TO 0.035"; 2) DECREASED THE DEPTH OF THE COUNTER BORE FROM 0.055" TO 0.050"; 3) INCREASED THE TOLERANCE FOR THE 0.15" DIMENSION OF THE TUBING POSITION FROM +-0.01 TO +-0.02; AND 4) ADDED A REQUIREMENT TO THE PRODUCT SPECIFICATION FOR SIDE LOAD FORCES: CONNECTOR MUST NOT OCCLUDE OR LEAK WHEN PRESSURIZED TO 1.0 +- 0.1 PSIG AIR PRESSURE AND A FORCE OF 4.8 +- 0.1 LBF IS DISTRIBUTED OVER THE CONNECTOR SURFACE AT AN ANGLE OF 12 DEGREES +- 2 DEGREES.