- Device
- IDENTITY/IDENTITY ADX, INFINITY AND INTEGRITY FAMILIES OF PACEMAKERS
- Applicant
- ABBOTT MEDICAL
- PMA number
- P880086
- Supplement
- S101
- Product code
- DXY
- Generic name
- implantable pacemaker Pulse-generator
- Decision date
- 2004-11-22
- Decision code
- APPR
- Date received
- 2004-08-30
- Supplement type
- Real-Time Process
- Supplement reason
- Change Design/Components/Specifications/Material
- Approval order statement
- APPROVAL FOR THE FOLLOWING: 1) A FIRMWARE MODIFICATION TO THE IDENTITY/IDENTITY ADX FAMILY OF PACEMAKERS TO CORRECT A POTENTIAL CONDITION THAT WOULD RESULT IN HARDWARE BACK-UP VVI PACING; AND 2) A COMPONENT DESIGN CHANGE TO IDENTITY/IDENTITY ADX, INFINITY AND INTEGRITY FAMILY OF PACEMAKERS MODIFYING THE PRINTED CIRCUIT BOARD USED TO CONNECT THE FEEDTHROUGH TO THE PACEMAKER CIRCUITRY.