Recall Z-2212-2026
- Recall number
- Z-2212-2026
- Event number
- 98820
- Firm
- KAYSERBETTEN GMBH & CO. KG Rieper Str. 12 Bad Fallingbostel Germany
- Firm FEI
- 3006795806
- Product code
- FNL
- Status
- Open, Classified
- Initiated
- 2026-05-04
- Posted
- 2026-06-04
- Root cause
- Device Design
- 510(k) numbers
- K915437, K103536, K900151, K972115, K946149, K911893, K963622, K974254, K963589, K854832, K881965, K881782, K926435, K951161, K881963, K872923, K790049, K811897, K875037, K830559, K962250, K864172, K873291, K951514, K881844, K852674, K872925, K930234, K894296, K903904
Product#
Pediatric care bed; Product Designation: KayserBett IDA;
Reason for Recall#
If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap themselves or others under the bed frame or between the bed frame and the floor. This can lead to serious injuries or even death of the patient (child) or third parties (children).