Recall Z-2212-2026

Recall number
Z-2212-2026
Event number
98820
Firm
KAYSERBETTEN GMBH & CO. KG Rieper Str. 12 Bad Fallingbostel Germany
Firm FEI
3006795806
Product code
FNL
Status
Open, Classified
Initiated
2026-05-04
Posted
2026-06-04
Root cause
Device Design
510(k) numbers
K915437, K103536, K900151, K972115, K946149, K911893, K963622, K974254, K963589, K854832, K881965, K881782, K926435, K951161, K881963, K872923, K790049, K811897, K875037, K830559, K962250, K864172, K873291, K951514, K881844, K852674, K872925, K930234, K894296, K903904

Product#

Pediatric care bed; Product Designation: KayserBett IDA;

Reason for Recall#

If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap themselves or others under the bed frame or between the bed frame and the floor. This can lead to serious injuries or even death of the patient (child) or third parties (children).