Recall Z-2310-2026

Recall number
Z-2310-2026
Event number
98569
Firm
Intuitive Surgical, Inc.
Firm FEI
3001675293
Product code
NAY
Status
Open, Classified
Initiated
2026-04-24
Posted
2026-06-04
Root cause
Device Design
510(k) numbers
K232610, K143217, K123329, K191529, K222839, K182140, K212101, K230033, K162973, K231798, K173415, K120215, K251213, K123463, K161178, K220023, K202166, K213710, K191736, K202968, K212054, K173906, K072627, K182371, K220889, K203632, K170875, K240852, K202834, K090993

Product#

Brand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/A Product Description: Preface: This report under 21 CFR 806.10 is being submitted for the dV5 system tower, as the NIR Handheld Camera or SOCK is developed and integrated as part of the dV5 system. The affected component of the dV5 system tower is the NIR handheld camera, and is supplied by Schoelly as the manufacturer. Therefore, the product description in this report references the NIR handheld camera system. Product Description: The NIR handheld camera system is compatible for use with the da Vinci 5 system to aid in surgery. The components of the NIR handheld camera system include a camera control unit and a light source, NIR handheld camera, and additional components that together, connect to the Tower of the da Vinci 5 system. When setting up the da Vinci 5 system, the image from the NIR handheld camera system is displayed on the Tower monitor. Component: Yes, the SOCK is the component and integrated into the dV5 system

Reason for Recall#

Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the handheld camera to the robotic endoscope.