Recall Z-2353-2026

Recall number
Z-2353-2026
Event number
98924
Firm FEI
3005994236
Product code
GWQ
Status
Open, Classified
Initiated
2026-05-04
Posted
2026-06-05
Root cause
Software Design Change
510(k) numbers
K171384, K221959, K071109, K791150, K212788, K113117, K781041, K791233, K162595, K791234, K910279, K964280, K010088, K903004, K854030, K033399, K180269, K181466, K231366, K970942, K051825, K002570, K010089, K870815, K010997, K081741, K191492, K863124, K921138, K904763

Product#

SD LTM 64 PLUS, EEG Amplifier/recorder, Model/Catalog Number: SD LTM 64 PLUS, Software Version: firmware 2021.02, 2022.01 or 2022.02

Reason for Recall#

Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.