Recall Z-2475-2026

Recall number
Z-2475-2026
Event number
98986
Firm
GE Medical Systems, LLC
Firm FEI
2126677
Product code
FMZ
Status
Open, Classified
Initiated
2026-05-01
Posted
2026-06-17
Root cause
Component design/selection
510(k) numbers
K020543, K101788, K070291, K870223, K102226, K101778, K071977, K020547, K061856, K993407, K182859, K821896, K162821, K813239, K801461, K943360, K061280, K993051, K914864, K152809, K102710, K810067, K132543, K912000, K003067, K870222, K813238, K072512, K913394, K974349

Product#

Giraffe OmniBed, a combination of an infant incubator and an infant warmer, All Models

Reason for Recall#

GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.