Recall Z-2482-2026
- Recall number
- Z-2482-2026
- Event number
- 98994
- Firm FEI
- 1417592
- Product code
- OHR
- Status
- Open, Classified
- Initiated
- 2026-05-04
- Posted
- 2026-06-17
- Root cause
- Nonconforming Material/Component
Product#
MEDLINE Urethral Catheterization Tray and Bag, Red Rubber Latex, 15 Fr, Medline Product Number/SKU (REF) DYND18350
Reason for Recall#
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.