Recall Z-2501-2026

Recall number
Z-2501-2026
Event number
98994
Firm FEI
1417592
Product code
MLS
Status
Open, Classified
Initiated
2026-05-04
Posted
2026-06-17
Root cause
Nonconforming Material/Component
510(k) numbers
K961993, K932216, K931722, K940518, K936120, K960243

Product#

MEDLINE Medical Procedure Kits labeled as: 1) BR L&D PACK, Medline Kit Number/SKU DYNJ47625B; 2) GYN VAGINAL, Medline Kit Number/SKU DYNJ912219; 3) KIT MAJ VAG SET UP, Medline Kit Number/SKU DYNJ907033B; 4) MINOR VAG PACK, Medline Kit Number/SKU DYNJ68663B; 5) OB DELIVERY PACK, Medline Kit Number/SKU DYNJ14099F; 6) OB DELIVERY PACK, Medline Kit Number/SKU DYNJ14099G; 7) OB PACK, Medline Kit Number/SKU DYNJ03387B; 8) OB PACK, Medline Kit Number/SKU DYNJ03387C; 9) OB PACK, Medline Kit Number/SKU DYNJ53543B; 10) RR-D/E PACK, Medline Kit Number/SKU DYNJ62645D; 11) RR-D/E PACK, Medline Kit Number/SKU DYNJ62645F; 12) RR-D/E PACK, Medline Kit Number/SKU DYNJ62645G; 13) VAGINAL DELIVERY PACK, Medline Kit Number/SKU DYNJ04454K; 14) VAGINAL DELIVERY PACK, Medline Kit Number/SKU DYNJ04454L; 15) VAGINAL PACK, Medline Kit Number/SKU DYNJ01830G.

Reason for Recall#

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.